Faster qualification process – Technology Transfer Package

The process of qualifying a shipping solution can be lengthy and take plenty of man hours. With the Technology Transfer Package Envirotainer hopes to relieve the customer in this process.

 

This service provides you with information required to qualify the Envirotainer containers for temperature sensitive product shipments. The information is mainly needed during qualification but also to collect knowledge on use and operation of the container.

The package includes material such as:

  • Operation Manuals
  • Specifications
  • Test reports
  • Maintenance information
  • Handling information

By simplifying the collection of documentation the customer can lower qualifying costs and easier adopt Envirotainer solutions.

The Technology Transfer Package is available online. If you wish to have access to the Technology Transfer Package, please contact your regional sales manager or send an email to Qualification@envirotainer.com.

Technology Transfer Package sign in

 

Envirotainer Cold Chain laboratory

The Envirotainer Cold Chain Laboratory consists of a dedicated team and two climate chambers with the possibility to simulate different time, temperature and humidity profiles. The chambers are located at our container manufacturing site in Rosersberg, Sweden. They were installed in 2014 by Weiss Umwelttechnik. For specifications and picture, please see below.

The climate chambers are mainly used for verification testing of containers in our development and product care process. In addition, data from container testing in the chambers is created to be available in our Technology Transfer Package (TTP) as a support in the qualification process of our containers.

In relevant cases the climate chambers may be used for testing on behalf of you as a customer. 

Please contact CCServices@envirotainer.com for any questions regarding TTP or Cold Chain Laboratory testing. 

 

Technical specifications

 

  WT 105’ (large chamber) WK 58’ (small chamber)
Size

105 m3
2 RAPs or 4 RKNs

58 m3
1 RAP or 2 RKNs
Temperature range -30 °C to +60 °C -30 °C to +60 °C
 

Humidity range

N/A

20% to 95% RH in +20 °C to +60 °C (empty chamber)

Temperature sensors 48 thermocouples type T 32 thermocouples type T
Chamber control system software SIMPATI SIMPATI
Data logging system software LabView LabView

 

The reference thermometer, used to calibrate the thermocouple sensors, is calibrated, according to standard ISO/IEC 17025, at an external laboratory accredited by a Swedish authority.

Cold Chain Laboratory Chamber WK 58’

Cold Chain Laboratory Chamber WK 58’